Main menu

Pages

GENERAL LAB



between peak broadening at low flow rates from diffusion and

at high flow rates from increased mass transfer effects between
mobile and stationary phases. It also predicts that, for a particular
chromatographic system, an optimal flow rate yields
the best efficiency. Empirically, the effects of flow rate on
efficiency are tested by measuring peak widths at different
flow rates. In practice, there usually is no interest in defining
the complete van Deemter curve, but rather in identifying the


highest flow


 rate that does not result in excessive peak broadening.
The usual goal is to identify conditions for the most
rapid analysis time that still yields adequate separations.
Factors that affect chromatographic efficiency are listed
in Box 13-1. In practice, efficiency is improved by using
(1) smaller particles, (2) nonporous particles, or (3) a coated
capillary surface as the stationary phase to minimize mass
transfer effects (the C constant in the van

 Deemter


 equation).
This allows the use of higher mobile phase flow rates without
peak broadening,

 resulting in faster separations. 

The limiting

factor in reducing particle size is the increased resistance to
flow and the need to operate at higher 
pressures.









You are now in the first article

Comments

The content